Sign in →

Test Code SGSS Synthetic Glucocorticoid Screen, Serum

Reporting Name

Synthetic Glucocorticoid Screen, S

Useful For

Confirming the presence of listed synthetic glucocorticoids in serum specimens (see Interpretation)

 

Confirming the cause of glucocorticoid-induced adrenal insufficiency

 

This test is not useful for detection of fluticasone propionate.

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Specimen Type

Serum


Ordering Guidance


 



Specimen Required


Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube:

Preferred: Red top

Acceptable: Serum gel

Submission Container/Tube: Plastic vial

Specimen Volume: 2 mL Serum

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

Serum: 1.1 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Frozen (preferred) 14 days
  Refrigerated  7 days
  Ambient  24 hours

Reference Values

Negative

Cutoff concentrations

Betamethasone: 0.10 mcg/dL

Budesonide: 0.20 mcg/dL

Dexamethasone: 0.10 mcg/dL

Fludrocortisone: 0.10 mcg/dL

Megestrol acetate: 0.10 mcg/dL

Methylprednisolone: 0.10 mcg/dL

Prednisolone: 0.10 mcg/dL

Prednisone: 0.10 mcg/dL

Triamcinolone acetonide: 0.10 mcg/dL

 

Values for normal patients not taking these synthetic glucocorticoids should be less than the cutoff concentration (detection limit).

Day(s) Performed

Wednesday

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

80299

LOINC Code Information

Test ID Test Order Name Order LOINC Value
SGSS Synthetic Glucocorticoid Screen, S 43141-1

 

Result ID Test Result Name Result LOINC Value
23593 Betamethasone 41745-1
23594 Budesonide 41747-7
23595 Dexamethasone 14062-4
23596 Fludrocortisone 41754-3
23600 Megestrol Acetate 41762-6
23601 Methylprednisolone 14186-1
23602 Prednisolone 12727-4
23603 Prednisone 12434-7
23605 Triamcinolone Acetonide 41767-5

Clinical Information

The primary use of this test is to assess exposure to synthetic glucocorticoids in patients with suspected central adrenal insufficiency.

 

Synthetic glucocorticoids are widely used as anti-inflammatory and immunosuppressive agents. Synthetic glucocorticoids can be used through various routes: oral, intravenous, intramuscular, inhaled, intranasal, and topical. In most cases, exposure to synthetic glucocorticoids is clearly documented in the medical record. However, occasionally, exposure to synthetic glucocorticoids is not clearly evident. This may occur when patients use non-prescription preparations (usually oral or topical) that have glucocorticoids.

 

Exposure to synthetic glucocorticoids leads to development of features of Cushing syndrome (weight gain, abdominal obesity, moon facies, skin thinning, easy bruising, and metabolic comorbidities) or to glucocorticoid-induced adrenal insufficiency due to hypothalamic-pituitary-adrenal axis suppression by the circulating supraphysiologic synthetic glucocorticoids. Hormonal work up in these cases demonstrates very low corticotropin and cortisol concentrations.

 

This test allows for measurement of nine synthetic glucocorticoids in patients in whom exogenous glucocorticoid exposure is suspected.

Interpretation

This test screens for and quantitates, if present, the following synthetic glucocorticoids: betamethasone, budesonide, dexamethasone, fludrocortisone, megestrol acetate, methylprednisolone, prednisolone, prednisone, and triamcinolone acetonide.

 

The presence of synthetic glucocorticoids in serum indicates current or recent use of these compounds. Since several of these compounds exceed the potency of endogenous cortisol by 1 or more orders of magnitude, even trace levels may cause hypothalamic-pituitary-adrenal axis suppression with development of glucocorticoid-induced adrenal insufficiency.

Cautions

This method cannot detect all available synthetic steroids available either as pharmaceutical compounds or chemicals present in food. The assay confirms only the listed synthetic glucocorticoids. For more information see Interpretation.

 

Lack of detection of listed synthetic glucocorticoids does not exclude previous exposure to one or more of tested glucocorticoids, as glucocorticoid-induced adrenal insufficiency (and resolution of features of overt Cushing syndrome) may persist for weeks-months after exogenous glucocorticoids were discontinued. In a patient with documented glucocorticoid exposure, circulating concentrations of synthetic glucocorticoids depend on the route of administration (detected for longer time if joint injection), dose and type of glucocorticoid that influence metabolism and clearance of the glucocorticoid (longer for triamcinolone).

Report Available

5 to 14 days

Reject Due To

Gross hemolysis OK
Gross lipemia OK
Gross icterus OK

Method Name

Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)